           1) Assemble

          a) Assemble cartridge adapters on the vacuum manifold

          b) Place ECUNIOGXF cartridge(s) on each vacuum station needed

          c) Connect the manifold to a suitable trap and attach the trap to a vacuum system capable of

          attaining a minimum of 25” Hg (635 mm) of vacuum

          d) Insert waste collection vial in manifold (optional)

          2) Prepare Water Sample

          a) Adjust the pH of the sample to less than 2 by adding 5 mL of 6N HCl or 2.5 mL of concentrated

          H2SO4

          b) If acid was added to the sample in the field do not add more unless the pH is greater than 2

          Gloves are recommended as skin oils may affect final sample weight

          3) Condition the Cartridge

          a) Rinse sides of cartridge and bottle holder with 10 mL of hexane

          b) Allow cartridge to soak for 1 minute

          c) Draw the hexane through the cartridge using vacuum

          d) Discard the hexane from waste vial

          e) Draw full vacuum through the cartridge for 2 minutes to dry

          f) Add 10 mL of methanol to the cartridge

          g) Slowly draw the methanol through leaving a layer on the cartridge frit

          h) Soak for one minute

          i) Remove waste vial and discard methanol

          From this point on do not let the cartridge dry out otherwise repeat at step f)

          j) Add 30 mL of DI water to the cartridge

          k) Draw the water through the cartridge to waste

          4) Sample Addition

          a) Add the 1 liter water sample directly to the cartridge

          b) Draw the sample through the cartridge under low vacuum. This may take several minutes

          depending on the solids in the sample. (Note 1) Increase vacuum pressure if necessary. Do

          not exceed 500 mL/minute

          c) Remove the cartridge and tap any excess water from the bottom of the cartridge

          d) Replace cartridge and allow to dry under full vacuum for 10 minutes

          e) Remove any water remaining in the bottom support of the cartridge with a paper towel if

          necessary

          5) Elution

          a) Prepare an extract collection vial containing about 8 mm (0.3 inch) of anhydrous sodium

          sulfate

          b) Place the vial in the manifold station under the cartridge

          c) Rinse the water sample bottle with 10 mL of hexane

          d) Add the hexane to the cartridge

          e) Soak cartridge for 2 minutes

          f) Turn on vacuum and slowly draw the hexane through the cartridge and into the collection vial

          g) Turn off vacuum then repeat steps 5 c) – f) 2 additional times but use 10 mL of hexane

          h) Do not allow the solvent to splash into the collection vial

          i) Add another 10 mL of hexane to the cartridge, rinsing the bottle holder

          j) Soak cartridge for 2 minutes

          k) Draw the hexane through the cartridge

          6) Dry the Extract

          a) Remove the collection vial from the manifold and cover with a screw cap

          b) Shake the extract to form a water/hexane emulsion and immediay pour the extract through a

          sodium sulfate funnel or column containing approximay 40 g of anhydrous sodium sulfate

          held in place with a glass wool plug or frit. Do not use filter paper

          c) Collect the extract in a clean, pre-weighed vessel

          d) Rinse the collection vial with hexane and add it to the sodium sulfate. This will rinse the vial

          and the sodium sulfatez

          7) Gravimetric analysis

          a) Carefully evaporate the hexane using a TurboVap® or similar evaporator at 40° C until the

          extract just reaches dryness.

          Note: Do not over dry or low recoveries will result

          Do not dry on a hot plate or in an oven or over dry as low recoveries may result

          b) Allow to cool to room temperature in a desiccator before weighing

          c) Record this weight as the mass per unit volume of oil and grease and report as HEM as mg/L

          Notes

          1) If very high solids are present, add glass wool to the cartridge prior to extraction to prevent clogging and improve

          flow. The glass wool must be thoroughly rinsed as part of the cartridge during the elution step.

          2) Stearic acid must be in solution in the spiking solution or low recoveries will result. If small crystals are present in

          the spiking solution, sonicate or shake until dissolved.

          DCN-118170-


          文件下載    
          廣州艾欣科學儀器有限公司
          地址:廣東省廣州市天河區(qū)元崗街道元崗路天河慧通產(chǎn)業(yè)廣場A1棟六樓1604(總部)
          郵箱:fighter31668@163.com
          傳真:86-020-36536169
          關注我們
          歡迎您關注我們的微信公眾號了解更多信息:
          歡迎您關注我們的微信公眾號
          了解更多信息
          主站蜘蛛池模板: 夜夜高潮夜夜爽夜夜爱爱一区| 色妞WW精品视频7777| 成人免费黄色网址| 亚洲国产成人久久综合一区77| 国产高清小视频| 天天碰免费上传视频| 亚洲欧美国产五月天综合| 黄色成人在线网站| 护士系列sdde221取精 | 久9久9精品视频在线观看| 精品一区精品二区制服| 天天摸天天碰天天爽天天弄 | 亚洲av无码第一区二区三区| 韩国三级香港三级日本三级| 国内揄拍高清国内精品对白| 九色综合九色综合色鬼| 美女被羞羞在线观看漫画| 天天爽夜夜爽人人爽一区二区| 久久亚洲精品成人777大小说| 男攻在开会男受在桌子底下| 国产精品成人免费视频网站| 久久午夜无码鲁丝片直播午夜精品| 精品一区二区久久| 国产免费私拍一区二区三区| zztt668.su黑料不打烊| 欧美啊v在线观看| 免费在线你懂的| 国产人与动zozo| 在线观看麻豆精品国产不卡| 中国特黄一级片| 欧美日韩第二页| 国产亚洲精品aaaaaaa片| wwwjizzz| 欧美va在线播放免费观看| 四虎精品视频在线永久免费观看| 99热这里只有精品免费播放 | 2021年北京马三1死2重伤| 日本处888xxxx| 亚洲人成无码www久久久| 熟妇人妻无码XXX视频| 国产小视频在线免费|